Intra-Aortic Balloon Pump Deflation, Auto-Triggering, Flow-Triggered Mechanical Ventilator
Gupta D, Duke JC, Rubens A, Orlewicz M. Intra-Aortic Balloon Pump Deflation Generated Suction-like Forward Aortic Blood Flow and Auto-Triggering of a Flow-Triggered Mechanical Ventilator. WebmedCentral CRITICAL CARE 2012;3(1):WMC002854
doi:
10.9754/journal.wmc.2012.002854
This is an open-access article distributed under the terms of the Creative Commons Attribution License(CC-BY), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
No
Abstract
Auto-triggering with the mechanical ventilators can be a problematic nuisance; however, it can be catastrophic too if the resultant respiratory alkalosis is not recognized early and treated with technical changes within the ventilatory mode and/or trigger settings of mechanical ventilation. We hereby present a postoperative case of a patient in which intra-aortic balloon pump (IABP) inflation-deflation cycle was triggering the mechanical ventilator. Intra-aortic balloon pump deflation generated suction-like forward aortic blood flow related auto-triggering of a flow-triggered mechanical ventilator should always be crossed off the checklist to correct the unexplained respiratory alkalosis in intensive care unit patients with cardiogenic shock who are either medically paralyzed or are on drug-induced paralysis (neuromuscular blockade).
Introduction
Auto-triggering with the mechanical ventilators can be a problematic nuisance; however, it can be catastrophic too if the resultant respiratory alkalosis is not recognized early and treated with technical changes within the ventilatory mode and/or trigger settings of mechanical ventilation. We hereby present a postoperative case of a patient in which intra-aortic balloon pump (IABP) inflation-deflation cycle was triggering the mechanical ventilator.
Case Report(s)
A 56 year-old-female was admitted to the surgical intensive care unit with open chest after aortic valve replacement for severe aortic stenosis and IABP in-situ for cardiogenic shock with severe pulmonary hypertension. Mechanical ventilation was maintained with assist/control mode at set respiratory rate of 14 breaths per minute with flow trigger at 2 liters/min. Neuromuscular blockade was maintained with continuous infusion of cisatracurium and titrated to the train-of-four twitches 1/4. On post-operative day 2, it was noticed that breath frequency has increased to 26 breaths per minute. Considering that patient might be breathing over the ventilator, the patient received a loading dose of 5-10 mg of cisatracurium to intensify the neuromuscular block to the train-of-four twitches 0/4. Arterial blood gases were drawn that showed respiratory alkalosis with pH 7.56. However, despite the additional dose of paralytic agent, the breath frequency did not improve. After re-consultation and re-evaluation with respiratory therapist who had observed similar case scenarios wherein patients with IABP in-situ and adequate neuromuscular blockade tended to hyperventilate over the set-ventilatory rate secondary to possible auto-triggering by the IABP, it was decided to change the ventilatory trigger to pressure trigger of 2 cm of water that resolved the hyperventilation and potentially catastrophic rising blood pH.
Discussion
The underlying mechanism for this auto-triggering was theorized to be secondary false positive flow trigger response to the forward blood flow in the thoracic descending aorta that happens due to the suction-like mechanism occurring due to the deflation of the IABP balloon at the onset of the systole to reduce the afterload. The suction-like phenomenon decompresses/empties the cardiac blood volume and increases the displaced pulmonary volume that eventually drops the expiratory flow that auto-cycles the flow-triggered mechanical ventilator to deliver non-initiated inspired breath to the patient. However, as the intrathoracic and pulmonary displacement changes is only limited to the flow, the auto-triggering is not initiated secondary to the pressure trigger that is dependent on generation of the negative intrathoracic pressure.
IABPs are commonly used to help increase myocardial perfusion in patients experiencing cardiogenic shock, intractable angina, low cardiac output after cardiopulmonary bypass, and as a bridge to further therapy in patients with MI, heart failure and arrhythmias to name a few (1). The pump is timed to inflate immediately after aortic valve closure thereby forcing blood into the coronary arteries, which branch off the proximal aorta, leading to increased myocardial perfusion. The IABP then deflates immediately before aortic valve opening, thus allowing forward flow of blood during systole (1). A search of the literature found a case report of autocycling in a patient with both and IABP and LVAD in use; however, the authors attributed the cause of the autocycling to the intrathoracic pressure gradients caused by the LVAD and not the IABP (2). The authors of that case reported that their patient’s hyperventilation was managed by turning off the trigger sensitivity and switching to SIMV. No documented case of isolated IABP-associated autocycling with flow trigger could be found. In another study, autocycling was shown to occur in 50% of patients with total artificial hearts (TAH’s) when using the flow-trigger setting (3). This was attributed to the significant pulmonary volume displacement caused by TAH oscillations while employing PEEP, which will work to maintain end-expiratory pressure at the expense of altered flow and, subsequently, may result in autocycling of the ventilator (3). We believe it is possible that the IABP device in our case caused the same pulmonary volume displacement as has been seen with TAH devices (4) and, therefore, lead to the autocycling that we observed.
Conclusion
In summary, intra-aortic balloon pump deflation generated suction-like forward aortic blood flow related auto-triggering of a flow-triggered mechanical ventilator should always be crossed off the checklist to correct the unexplained respiratory alkalosis in intensive care unit patients with cardiogenic shock who are either medically or surgically paralyzed or are on drug-induced paralysis (neuromuscular blockade).
References
1.http://www.uptodate.com/contents/intraaortic-balloon-pump-counterpulsation?source=search_result &search=iabp&selectedTitle=1~58#H15 Last accessed on December 26, 2011.
2.http://www.rcjournal.com/abstracts/2009/?id=679097 Last accesssed on December 26, 2011.
3.Shoham AB, Patel B, Arabia FA, Murray MJ. Mechanical ventilation and the total artificial heart: optimal ventilator trigger to avoid post-operative autocycling - a case series and literature review. J Cardiothorac Surg. 2010 May 17;5:39.
4.Rouby JJ, Leger P, Arthaud M, Devilliers C, Cabrol A, Gandjbakch I, Cabrol C, Viars P. Respiratory effects of the Jarvik-7 artificial heart. J Appl Physiol. 1989 Apr;66(4):1984-9.
Source(s) of Funding
None
Competing Interests
None
Disclaimer
This article has been downloaded from WebmedCentral. With our unique author driven post publication peer
review, contents posted on this web portal do not undergo any prepublication peer or editorial review. It is
completely the responsibility of the authors to ensure not only scientific and ethical standards of the manuscript
but also its grammatical accuracy. Authors must ensure that they obtain all the necessary permissions before
submitting any information that requires obtaining a consent or approval from a third party. Authors should also
ensure not to submit any information which they do not have the copyright of or of which they have transferred
the copyrights to a third party.
Contents on WebmedCentral are purely for biomedical researchers and scientists. They are not meant to cater to
the needs of an individual patient. The web portal or any content(s) therein is neither designed to support, nor
replace, the relationship that exists between a patient/site visitor and his/her physician. Your use of the
WebmedCentral site and its contents is entirely at your own risk. We do not take any responsibility for any harm
that you may suffer or inflict on a third person by following the contents of this website.